68
[
α
]Q
Same as the operation of pressing the
key of the balance "
α
."
[
α
Same as the operation of pressing the
key of the balance "
α
."
[
α
]D05
Same as the operation of pressing the
key of the balance "
α
."
[
α
]D07
Same as sending the D07 command to the balance "
α
."
[
α
]UNIT
Same as the operation of pressing the
key of the balance "
α
."
[
α
]RANGE
Same as the operation of pressing the
key of the balance "
α
."(Valid
only for D-type)
[
α
]CAL
Same operation as sending the CAL command to the balance "
α
."
[
α
]UNIT+
Same as the state of holding down the
key of the balance "
α
."
Format of Output Data from the Balance
The output data from the balance "
α
" is performed with the following format.
"[
α
]" data <delimiter>
Format of the load data is the same as the normal output form (
).
Operation is not guaranteed for the data containing characters, such as printing of date, time, and
calibration document etc.
Restricted Items in the Multi-Connection Mode
Multi-Connection mode is not designed for each balance to independently send the data.
This mode is for sampling the data by control of multiple balances with one PC. This is not the
function to support multiple balances for sending the data separately.
When multiple number of data and commands exist on the system at the same time, the balance may
not operate normally. For example, multiple data may exist simultaneously on the system when using
multiple balances in the Application Measurement mode such as Auto Print or Continuous Output, or
pressing the
key. Normal operation cannot be guaranteed.
Peripheral instruments connected to the DATA I/O or IFB-102A connector, such as an EP-50
Electronic Printer, cannot be used.
The communication formats can only be used with
. Delimiters cannot be used with
or
.
It will take an average 0.05 x N seconds for the data from the personal computer to reach balance unit
N, even at setting of 38400 bps. Also, it will take almost the same time for the data from the balance
in the Nth position to reach the personal computer.
Cancellation of Multi-Connection mode
Multi-Connection mode cannot be canceled unless the power for all balances is turned OFF.
Summary of Contents for CBW
Page 1: ...2 CBX 앞표지 ...
Page 2: ...3 PRECAUTIONS ...
Page 3: ...4 ...
Page 27: ...28 ...
Page 79: ...80 MEMO ...
Page 80: ...81 MEMO ...
Page 81: ...82 MEMO ...
Page 82: ...83 ...
Page 83: ...84 ...