37
2.5.5 Calibration Check Using External Weights
1)
Verify that the balance is in mass display and unload the sample from the pan.
2)
Press the
key once to display
.
(If
is not displayed, select Menu Map item [4].)
3)
Press the
key.
The value of the correct calibration mass to be loaded is displayed and blinks.
Changing the Calibration Mass to be Used
Pressing the
key allows changes to the weight value. Modify the value
using the
key and
key, then press the
key. To
interrupt modification, press the
key.
4)
Load the indicated calibration mass and press the
key.
The zero display blinks.
5)
Unload the weight from the pan and press the
key.
The display changes to the
xxx display. (xxx indicates a numeric value)
6)
To change the
value to zero, press the
key.
Press the
key to avoid changing the value to zero.
Pressing the
key interrupts calibration and does not change this value
to zero.
Note:
Changing the value to zero is equivalent to performing span calibration.