2.10 Unit Registration Menu and Unit Change
:
2.10.1
Unit Registration Menu
Press the
key in mass display to sequentially change the registered units. It is possible to
display units other than "kg" with the CBX-K and CBW-K series of balances. Before weighing, set the
display units to be used with the unit setting menu [54] to [62]. (Unit kg, %, number (PCS) were set
before shipment.)
Note:
•
In the unit setting menu, the stability symbol is illuminated to indicate the currently set units.
•
Set or release the unit by pressing the
key when the unit is displayed.
Functional Units:
Solid specific gravity unit (
▼
d):
Density of the liquid in which the sample is immersed.
Liquid specific gravity unit (d):
Volume of the reference weight to be immersed in the sample.
unit (
►
):
Numeric value (multiplier) multiplied by the gram (g)
weight of the unit.
To register Functional Units, set the constant to something
other than 0. Setting the constant to 0 turns the function off.
For the unit names, refer to
Section 2.3 "Menu Map."
2.10.2 % Conversion
1) Set the % unit with Menu Map item [56].
The % unit is set before shipment.
2) Press the
key several times in the mass display until the % unit is displayed.
Setting the 100% reference
Press the
key to tare the balance.
Load the reference sample that corresponds to the 100%value. This value must be equivalent to 100
counts or more in the "kg" unit.
When the stability symbol illuminates, press the
key.
is displayed briefly and the reference sample weight is displayed as 100%.