3.
Now you’ll see the screen shown in Figure 7, which allows you to select the
channel (that is, the specific 2016) that is to be calibrated.
4.
Press OK to enable the selected channel, default enable the first SCA on the
selected channel, display the correct detector datasource in an MCA View
Control window, reset the calibration information for the detector input, clear
it and start acquisition. Now the screen shown in Figure 8 will be displayed.
Adjust the selected 2016’s gain until the energies of interest are displayed in
the selected SCA windows (Figure 8). The RDY LED on the selected module
will start blinking.
13
Calibration Procedure
Figure 6 Starting the Calibration Procedure
Figure 7 Selecting the 2016 to Calibrate