EASY-DRIVER – User’s Manual
18
2.5
Load connection
Turn off the AC input power before making or changing any
rear panel connection, Ensure that all connections are
securely tightened before applying power. There is a
potential shock hazard when using a power supply with a
rated output greater than 60 V
Wire selection
Two factors must be considered for the selection of the wires:
•
Current carrying capacity -> Cross section area
•
Maximum wire length.
•
Insulation voltage
Wire cross section and length
Wire size should be selected to enable voltage drop per lead to be less than 1 V
at the maximum power supply current to prevent excessive output power consumption.
Suggested wire sizes are listed in the following table:
Wire Cross Section Area [mm2]
Resistivity [Ω/km]
Maximum cable length in meters to
limit voltage drop to be less than 2 V
(1 V per lead)
5 A
10 A
2.5
8.21
24
12
4
5.09
39
19.5
6
3.39
59
29.5
10
1.95
102.5
51
16
1.24
161
80.5
Table 2:
Wire selection – example for 0520 and 1020 models
Wire range allowed from the output connectors is from 0.5 to 16 mm
2
, but wire
cross section below 2.5 mm
2
are discourage to be used for minimize power losses and
voltage drops. If values of Table 2 are used the maximum voltage to the load will be
limited to:
EASY-DRIVER nominal output voltage – maximum 20 V
EASY-DRIVER power supplies could be connected to the load with a single
insulation wires which have voltage rating of at least 60 V
DC
.