EASY-DRIVER User’s Manual
55
7.
EASY-DRIVER Remote Reboot
The EASY-DRIVER module can be rebooted either locally (by pushing the
RESET button on the front panel) or remotely.
The procedure to remotely reboot the module is the following:
A.
Establish a UDP connection with the EASY-DRIVER on the
port 30704
B.
write to the module port the following 9-bytes commands, separated with
100ms waits:
o
0x 19 04 00 00 00 04 00 00 00
o
wait 100ms
o
0x 1A 04 00 00 00 00 00 00 00
o
wait 100ms
o
0x 1B 04 00 00 00 00 00 00 00
o
wait 100ms
o
0x 1B 04 00 00 00 04 00 00 00
C.
now the module should turn on again and will be ready for operation in about
5s.
It is important to notice that after executing point C., connection to
port 30704 must be closed and, in order to correctly operate the power
supply again, a new connection to the EASY-DRIVER module IP address
standard port - i.e. 10001 - must be established.
The reboot process here describe will not reboot the Ethernet to Serial converter
(XPort). Reboot of the XPort device can be done only by a power cycle, i.e. switching
Off and On the main switch.