EASY-DRIVER User’s Manual
53
6.1
DeviceInstaller® Software
The DeviceInstaller® software can be downloaded from the Lantronix website
www.lantronix.com
and the Telnet connection can be established either using the
software itself or a simple command line socket (on all operating systems). The EASY-
DRIVER power supplies can be connected to a global LAN or point-to-point
(recommended in order to obtain minimum delays, maximum speed performance and
to avoid possible communication problems). Please note that for a point-to-point direct
connection a twisted Ethernet cable must be used.
The next few steps must be followed in order to assign a new IP address to the
module:
•
Connect to the EASY-DRIVER unit with a twisted Ethernet cable;
•
Verify that the “Link LED” on the RJ45 connector is turned on (amber for a
10Mbps connection or green for a 100Mbps connection);
•
Launch the “DeviceInstaller” program;
•
Select the XPort device that you want to change the IP address to;
•
Click on the “Assign IP” icon;