l
Switch on the main disconnect switch, then release the emergency
stop button. Restart the pump by pressing on Start button on the
remote control. If the pump does not start, press on emergency
stop button, switch off and padlock the main disconnect switch,
switch off the customer’s power supply and put the label
“maintenance processing”. Call the BUSCH Semicon service team
in charge of the maintenance of the pumps.
l
If the external thermal protection B1 is engaged, check that the
water flow shows 2l/min on the flowmeter. If not, set the water
flow at 1,5 l/min according to the instruction (see chapter
“Checking the cooling liquid flow”). Check the wiring system of
the thermal protection by opening the side face (Type 3 of
electrical work). Wait a few hours for the cooling of the protection,
then press on Start button to restart the pump.
Safety interlock description
Interlock event
Sensor
Response
Emergency
Stop
EMO S/W
EMO Button Pushed during pump
running
H
Motor
overcurrent
Frequency
inverter
Inverter default and remove
power to motor
s
Motor overtem-
perature
(t° > 150°C)
Thermo
switch B1
(klixon)
Contactors open and remove
power to motor
s
Discharge
overpressure
Pressure
sensor B4
Depends on alarm function set in
the PLC for overpressure
Set at factory: General stop ie. the
contactors open and remove
power to motor
S
Motor
overcurrent
Frequency
inverter
Depends on alarm function set in
the PLC for overcurrent
Set at factory: General Stop ie. the
contactors open and remove
power to motor
S
Cylinder
overtemperature
Temperature
sensor B5
Depends on alarm function set in
the PLC for cylinder temperature
Set at factory: General Stop ie. the
contactors open and remove
power to motor
S
Cooling water
flow low
Water
flowmeter B2
Depends on alarm function set in
the PLC for water flow
Set at factory: General Stop ie. the
contactors open and remove
power to motor
S
Nitrogen flow
too low
Nitrogen
flowmeter B3
Depends on alarm function set in
the PLC for nitrogen flow
Set at factory: None ie. the pump
continues running with alarm
S
H= Hardware, S= Software
Position of the pump gravity center
l
The gravity center of the BC 0200 F pump is determined according
to the drawing
Installation of the vacuum pump in a
seismic zone
l
Convey the pump to its final location with a pallet truck before
removing it from its support
l
Unscrew the fixing screws of the pump on the transport pallet and
remove the machine from its support with an appropriate lifting
system by using the lifting brackets made for this purpose
l
Check that the wheels no longer touch the ground
CAUTION
The vacuum pump is fixed to the ground with four M10 screws, one
screw per bracket. The fixing screw must stand a 1200 N tensible
force (per bracket).
The final user must adapt the type of screws depending on the
ground material.
Information over the lubricants
Oil
Busch YLC 250 B n°.-art. 0831 131 400 (0.5l of Busch YLC 250 B)
Quantity: BC 0200 F : 0,12 l
Replacement: After 16'000 h (see Maintenance chapter, page 15)
Decommissioning procedure
When the product arrives at the end of the lifetime, it is necessary to
proceed to the decontamination of the vacuum pump
Safety
BC 0200 F
Page 8
0870767417 (En)
ELECTRICAL
POW ER
Main
switch S1
Frequency
inverter
DP Motor
Klixon
B1
ELECTRICAL PRINCIPLE OF OPERATION
Emergency
stop
Main
Circuit
Breaker
Q1
Wheel
Adjusting nut
Lifting bracket
Dimensions :
L1 : 264 [mm]
L2 : 132 [mm]
L3 : 497 [mm]
h : 97.7 [mm]
W : 130 kg
Center of gravity posi
Ɵ
on
CG : Center of gravity
W : Weight of vacuum pump