Test protocol of
an inspection of
electro-medical
equipment
according to
DIN EN 62353:
2015-10
Page 1 of 2
Customer / med. Facility / practice:
Address:
Carried out:
repeat inspection
inspection prior initial operation
_____________________________________________
inspection following repairs / servicing
Equipment type:
hospital bed
care bed
˛
Slatted frame
Protection classification:
i
˛
ii
Bed type:
LIPPE IV
Inventory Number:
Location:
Serial Number:
Grade level (IEC60601-2-52):
1
2
3
4
5
Manufacturer:
Burmeier GmbH & Co. KG
User-specific:
none
Testing equipment used (Type / Inventory Number):
1.
MPG Classification:
Class I
2.
I. Visual Inspection
OK
Not
OK
Defect Description
Visual inspection of the electriacl components
Stickers and type plates
Presence, legible
housing of power adapter unit / Power adapter
Damage
Motor housing and lifting rods of motors
correct position, Damage
hand switch
Damage, foil intact
Motor-, hand switch- and main cables
Damage, routing
Plug and plug covering strip of control unit
available, correct position
Visual inspection of the mechanical components
Stickers and type plates
Presence, legible
Safe working load / patient weight
Presence, legible
Trapeze pole/socket
Damage, Deformation
bed frame
Damage, Deformation
Spring wood slats
Damage, Splinters
castors
Damage
reclining surface
Damage, Deformation
w ooden boards
Damage, Splinters
w elded seams
Split welded seams
Side rail
Damage, Splinters
Knurled screws
Tight fit
w earing parts such as joints
Damage
II. Electric measurments according to DIN EN 62353:2015-10 (leakage current, direct)
Proceed as follow:
OK
Not
OK
Defect Description
1. Plug the power adapter into the test socket on the measuring instrument.
2. insert the plug of the 24 volt cable into the special adapter (extra accessory; request from bUr Meier ).
3. connect the special adapter with the test probe of the test unit.
4. Start procedures with the measuring instrument
Limit Value
Value of first
measurement
current actual
value
result: bed SK ii (T ype b)
0,1 ma
ma
ma