Operating instructions DuMaster D-480
©BÜCHI Labortechnik AG
4 - Understanding the instrument and planning its use
65
The statistics view displays the element content percentages of all samples in a series, grouped by
sample name.
For each group of samples, the following calculated statistical values per element are also displayed:
Percentage mean value of the element
The absolute standard deviation of the mean values
The relative standard deviation of the mean values
Difference (Delta) of the highest and lowest content of the group
You can exclude individual samples and/or values of a group from the calculation of statistical values
or include them again.
Calibration view
If your series includes samples previously defined as calibration samples, you can use them to
calculate calibration coefficients. You can define and determine up to two calibration ranges
according to which polynomial degree you want to determine the calibration curve.
In this case, the calibration view displays the following information:
For each defined calibration range a graph display of the calibration curve optimized for the
respective calibration samples.
The calibration coefficients computed according to the selected polynomial degree.
The correlation coefficient for the depicted calibration curve that allows you to assess the quality
of calibration.
Summary of Contents for DuMaster D-480
Page 1: ...DuMaster D 480 Operation Manual 11593626C en...
Page 4: ......
Page 12: ......
Page 16: ......
Page 30: ......
Page 102: ......
Page 181: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 8 Maintaining the instrument 181...
Page 338: ......
Page 352: ......
Page 353: ......