Operating instructions DuMaster D-480
©BÜCHI Labortechnik AG
2 - Basic security settings
28
Installing/removing the drying tube
Please observe the following instructions:
Caution
Lack of ventilation of the analyzer!
A lack of ventilation leads to component overheating
Do not open the front door too long when the furnaces are hot and the furnace is
not pulled out. Otherwise the ventilation will not work properly.
Warning
Caustic substance Sicapent® (phosphorus pentoxide)!
Danger of chemical burns when handling Sicapent®.
Wear protective glasses and the enclosed protective leather gloves.
Overheating of tube fillings
Please observe the following instruction:
Caution
Overheating if tube fillings are not appropriate for the operating mode!
Overheated tube fillings melt, run into the furnace area and destroy the furnace.
Make sure that the tube fillings correspond to the selected operating mode.
Working with the database
Please observe the following instructions:
Caution
Data loss
After reorganizing the database via
Clean
all sample data and the contents of the
logbook are irrevocably deleted.
Always backup your data before reorganizing the database via
Clean
.
Caution
Data loss
Restoring the database via
Restore
irrevocably overwrites all sample data, the
contents of the logbook and the database configfuration parameters.
Always backup your data before restoring the database via
Restore
.
Improper changes
Please observe the following instruction:
Caution
Improper changes impair proper operation of the system and may destroy system
components.
Only configure the instrument if you are authorized to do so and observe the
operating instructions.
Summary of Contents for DuMaster D-480
Page 1: ...DuMaster D 480 Operation Manual 11593626C en...
Page 4: ......
Page 12: ......
Page 16: ......
Page 30: ......
Page 102: ......
Page 181: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 8 Maintaining the instrument 181...
Page 338: ......
Page 352: ......
Page 353: ......