Operating instructions DuMaster D-480
©BÜCHI Labortechnik AG
8 - Maintaining the instrument
177
CalibrationLow (0 - 300 mg):
O
2
dosing time 90 sec.
O
2
flow = 225 ml/min.
O
2
cut-off threshold = 15 %.
CalibrationHigh (300 - 700 mg):
O
2
dosing time 150 sec.
O
2
flow = 300 ml/min.
O
2
cut-off threshold = 20 %.
Start the analysis process by selecting
System > Auto run
.
Check whether the O
2
field "MFC O2" on the screen displays the expected setpoint value during
the combustion time. See Values for oxygen dosing.
Observe the measuring results. If the N blank value is not below 50 after stabilization: Stop
continuous analysis via
System > Single
. Find out the reason of the increased blank value and
correct it. Cf. Types of blank value determination and their settings.
Save the original values. See
Data backup and printing
(on page 156).
Phase 2: Ascertaining appropriate calculation methods
Proceed as follows:
Execute the following command:
Math. > Calibration
. If sequential blank value determination is
activated a message window opens showing the corresponding information.
Click
Yes
to continue. The "Configure calibration" dialog appears.
Define the following settings:
Check the box next to "Two calibration ranges".
Enter 1 as the polynomial degree for the lower and upper range. This corresponds to linear
calculation.
Click
OK
to save your settings and close the dialog. The "Specify lower and upper calibration
range" dialog appears. It displays a graph generated from the absolute contents of the samples
and their peak areas in a linear calculation. The graph displays deviations of the measuring
points from these lines.
Observe the measuring points and decide
what calculation would be suitable
whether you would separate the the ranges
and, if so, where you would separate the ranges.
See
Calibration curve calculation method criteria
(on page 52).
Write down your considerations.
Click
OK
to view the readings for the other elements. For each element, decide as described
before.
Write down your considerations.
Click
Cancel
.
Phase 3: Set calculation methods, perform calculation
Proceed as follows:
Execute the following command again:
Math. > Calibration
. If sequential blank value
determination is activated a message window opens showing the corresponding information.
Click
Yes
to continue. The "Configure calibration" dialog appears.
Define the settings that you ascertained in the preceding phase.
Summary of Contents for DuMaster D-480
Page 1: ...DuMaster D 480 Operation Manual 11593626C en...
Page 4: ......
Page 12: ......
Page 16: ......
Page 30: ......
Page 102: ......
Page 181: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 8 Maintaining the instrument 181...
Page 338: ......
Page 352: ......
Page 353: ......