Operating instructions DuMaster D-480
©BÜCHI Labortechnik AG
4 - Understanding the instrument and planning its use
50
Instrument equipment
Combustion tube filling
The following picture shows the fillings and filling heights of the combustion tube:
Legend:
1 Ash crucible
2 Corundum balls (10 mm)
3 80 g CuO + 20 g Corundum balls (87 mm)
4 Corundum balls (26 mm)
5 Wire mesh ronde
6 Spacer
Filling the post combustion tube
The following picture shows the fillings and filling heights of the post combustion tube:
Legend:
1 Plug with embedded 3.3g brass wool
2 53g Regainer (126 mm)
3 Quartz wool (10 mm)
4,180 g CuO + 15 g Pt catalyst (167 mm)
5 Corundum balls (28 mm)
6 Wire mesh ronde
7 Spacer