Operating instructions DuMaster D-480
©BÜCHI Labortechnik AG
8 - Maintaining the instrument
209
Caution:
Tilting and jamming the combustion tube connection can damage the combustion tube.
Pull the combustion tube connection (3) out of the combustion tube with careful jiggling motions.
Place the combustion tube connection with Ni flap (3) on the tube rack.
Allow the flange to cool down to room temperature.
Phase 3: Cleaning the Ni flap
Proceed as follows:
Slowly pull out the O
2
lance.
Remove sooty particles from the metal surface by means of a wire brush. For cleaning the Ni flap
itself use a soft brush (e.g. toothbrush).
In case of heavy contamination, both sides must be cleaned. For this purpose remove the o-rings
and the O
2
lance.
Place the complete combustion tube connection with the Ni flap in ethyl alcohol or in a laboratory
cleaner (duration approx. 15 min).
Brush again the metal surface and the flap itself as described above.
Remove possible residues of ethyl alcohol with water and allow the combustion tube connection
to dry.
Insert the o-rings again.
Slightly grease the o-rings.
Insert the O
2
lance again.
Installing the Ni flap
Note
The following section is a continuation of "Removing the Ni flap". All steps described in "Removing the
Ni flap" must already have been performed. See: Removing and cleaning the Ni flap.
Summary of Contents for DuMaster D-480
Page 1: ...DuMaster D 480 Operation Manual 11593626C en...
Page 4: ......
Page 12: ......
Page 16: ......
Page 30: ......
Page 102: ......
Page 181: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 8 Maintaining the instrument 181...
Page 338: ......
Page 352: ......
Page 353: ......