Operating instructions DuMaster D-480
©BÜCHI Labortechnik AG
8 - Maintaining the instrument
221
Emptying reaction tubes
Requirements:
Before starting work, the following requirements must be met:
The reaction tubes have been removed from the furnace.
The reaction tubes have cooled down.
Keep a tube extractor ready.
Safety instruction
Strictly observe the safety instruction "Emptying the reaction tubes" under
Warning notes during
operation
(on page 24).
Emptying the combustion tube
Proceed as follows:
Insert the tube extractor into the combustion tube. After clamping the crucible tongs in the ash
crucible, pull the ash crucible out.
Empty the ash crucible.
Empty the copper oxide out of the combustion tube.
Remove the wire mesh ronde and the spacer out of the combustion tube.
Remove the CuO dust out of the combustion tube by means of a damp cloth.
Dispose of the chemicals according to the relevant disposal categories.
Emptying the post combustion tube
Proceed as follows:
Remove the brass wool out of the post combustion tube plug. Usually, the brass wool has turned
to discolored and is used-up.
Remove the Regainer.
Remove the corundum balls.
Remove the CuO and the Pt catalyst.
Remove the corundum balls.
Remove the wire mesh ronde and the spacer.
Remove the CuO dust and, if applicable, the crust of salt by means of a damp cloth.
Dispose of the chemicals according to the relevant disposal categories.
Dispose of the Reductor
Proceed as follows:
Dispose of the Reductor to the relevant disposal categories.
Filling the combustion tube
Note
Note the following rules:
Observe the filling heights indicated as exactly as possible.
Summary of Contents for DuMaster D-480
Page 1: ...DuMaster D 480 Operation Manual 11593626C en...
Page 4: ......
Page 12: ......
Page 16: ......
Page 30: ......
Page 102: ......
Page 181: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 8 Maintaining the instrument 181...
Page 338: ......
Page 352: ......
Page 353: ......