Operating instructions DuMaster D-480
©BÜCHI Labortechnik AG
3 - Product description
34
Dimensions and weights
The following table contains dimensions and weight of the instrument:
Reference value
Technical specifications
Dimensions
480 mm x 550 mm x 550 mm (Width x Depth x Height)
Weight
65 kg
Electrical and gas supply
The following table contains the technical specifications of the electrical and gas supply:
Reference value
Technical specifications
Mains voltage
100–230 Volt AC ± 10 %; 50–60 Hz
Power requirement
0.6 kWh/h
Connected load
1.4 kVA
Connected electrical loads and
frequency of the complete
analysis unit
Connected electrical load:
Voltage:
Frequency:
16 A
100V AC ± 10 %
120V AC ± 10 %
230V AC ± 10 %
50-60 Hz
Supply gases
Carbon dioxide:
Oxygen:
Purity: 99.995 %
Purity: 99.995 %
Consumption of supply gases
Per analysis approx. 2.5 liters carbon dioxide.
Per analysis approx. 0.5 liters oxygen.
Ambient conditions
The following table contains the technical specifications of the ambient conditions:
Reference value
Technical specifications
Protection class
Protection class I, protective conductor connection
Mode of protection
IP20, installation in dry rooms only
Noise level
55 dB (A), as per DIN 45635 section 1
Permissible ambient temperature
Maximum: + 35 °C, minimum: + 15 °C
Summary of Contents for DuMaster D-480
Page 1: ...DuMaster D 480 Operation Manual 11593626C en...
Page 4: ......
Page 12: ......
Page 16: ......
Page 30: ......
Page 102: ......
Page 181: ...Operating instructions DuMaster D 480 B CHI Labortechnik AG 8 Maintaining the instrument 181...
Page 338: ......
Page 352: ......
Page 353: ......