Operating instructions DuMaster D-480
©BÜCHI Labortechnik AG
2 - Basic security settings
29
Baking out combustion tube and reductor
Please observe the following instructions:
Warning
Gas pressure and caustic substances in the instrument!
Consumables may escape under pressure and cause chemical burns. Before
performing the work:
Shut off the gas supply. To do so, execute the
Options > Maintenance >
Replace parts
command.
Warning
Hot components in the instrument
When working inside the instrument there is a risk of burning as many parts of the
instrument are hot.
When working inside the instrument always wear protective glasses and the
enclosed heat protection gloves.
Mixing up gas inlets
Please observe the following instruction:
Caution
A mix-up of the gas inlets may lead to destruction of the instrument.
Connect the supply lines carefully and observe the diagram.