Option cards
8.9 Analog Scale Input Option with 5VDC Excitation
This card allows you to connect a second analog scale with 5VDC excitation.
Figure 8.12 5VDC Excitation Analog Scale Input option card
SENSE Excitation Jumpers
l
4 wire loadcells requre
jumpers to be ON the
pins.
l
6 wire loadcellss require
jumpers to be OFF the
pins.
TB1
Pin
Function
1
-EXC (0V)
2
+EXC (+5V)
3
-SEN
4
+SEN
5
-SIG
6
+SIG
7
SHLD
S1 switch settings must
all be OFF to operate.