32
T419S Indicator Service Manual
6
ADMIN Menu
6.1.10 DigJBox procedure
If you pick the
DigJBox
item, the screen shown in
Figure 6.5
appears. This refers to
the ZB210 digital junction box.
Figure 6.5 Calibration menu
The
Corner
menu item is for balancing the load cell outputs using weights placed over
each corner of the scale. See
.
The
Section
menu item is for balancing the load cell outputs using weights placed
between load cells or sections of the scale. See
Section Weight Adjustment on page
32
.
The
Swap
menu item is for replacing a bad load cell with a good one and still be
assured the accuracy remains high. See
Sensor Swap Procedure on page 33
.
Corner Balancing
Highlight
Corner
and press
Enter
. Follow all the on-screen prompts to complete the
cornering process. When completed you will be asked if you want to perform a Ghost
calibration. Answer accordingly and follow the instructions. Ghost calibration is not
legal for trade.
When finished, follow the Zero and Span procedures in your indicator’s service manual
to finish the calibration.
Section Weight Adjustment
For optimum section calibration, enter the Weigh Bar nominal span Factors and cell
capacities in ZTools before performing Section weight adjustments. If no span values
are available for the sensor, you MUST enter
1
for a value. Leaving it at
0
disables
section weight adjustment.
As long as you use the same weight the system will adjust the sections correctly.
Remember, a minimum of 10% scale capacity is recommended.
As long as you use the same weight over each sensor, the system will corner
correctly. Remember, a minimum of 10% scale capacity is recommended.
Summary of Contents for T419S
Page 2: ...BTEK_T419S_s_en_501788 book 2019 B TEK All rights reserved ...
Page 118: ...118 T419S Indicator Service Manual ...
Page 130: ...130 T419S Indicator Service Manual ...
Page 134: ...134 T419S Indicator Service Manual Last 28 Temp 28 ...
Page 135: ......
Page 136: ...B TEK Scale LLC 1510 Metric Ave SW Canton OH 44706 3088 ...