General
testing
conditions
Radiation measurements were conducted –
Using a Thermo micro-rem low energy plastic scintillation radiation survey instrument.
With the survey meter calibrated by the survey meter manufacturer.
Referenced to the center of the detector volume.
Readings
Readings were taken at 50kV, 39 µA, with a Duplex 2205 sample over the examination window
and rounded up to the nearest 5 µRem/hr value.
Displayed isodistance contours show the maximum value likely to be encountered in x10
-6
Rem/hr
(
rem/hr) or equivalently in x10
-8
Sieverts/hr.
Duplex
μrad(μSieverts)/hr
Location
10cm
30cm
100cm
A
≤45(0.45)
≤20(0.20)
≤5(0.05)
B
≤10(0.10)
Background
Background
C
≤5(0.05)
≤5(0.05)
Background
D
Background
Background
Background
E
Background
Background
Background
F
Background
Background
Background
G
Background
Background
Background
H
≤40(0.40)
≤5(0.05)
Background
I
≤40(0.40)
≤5(0.05)
Background
J
≤10(0.10)
Background
Background
K
≤10(0.10)
Background
Background
L
≤5(0.05)
Background
Background
M
≤5(0.05)
Background
Background