S1 TITAN, T
RACER
5, and CTX User Manual
030.0112.04.0
42 of 67
Phase length
To manually set the length of each phase –
Step
Action
Result
1
From the
Ready to Test
screen, tap
SETTINGS
.
The
SETTINGS
screen is displayed.
2
Ensure that
Use Default Settings
box is not
checked. If it checked, tap it.
Phase options are displayed.
3
Ensure the box below
Unlimited
is unchecked.
The box is empty and fields previously
grayed out are available.
4
Set each phase duration in seconds by tapping
either the minus sign (
-
) or the plus sign (
+
) to
decrease or increase, respectively, the
measurement duration.
OR
Tap the field.
The field reflects the number change.
OR
A numeric keyboard is displayed.
5
On the keyboard, tap a number and
OK
.
The keyboard is removed and the field
reflects the number change.
6
For subsequent phases, repeat steps 3-5.
To disable subsequent phases
, set duration
to 0.
7
Tap
OK
.
The Settings are saved and the
Ready
to Test
screen is displayed.
One phase
If a method does not use multiple phases, only one phase is displayed.
Unlimited
Unlimited
indicates the measurement continues while the trigger is pressed and ends when the
trigger is released. It is not a timed measurement but is limited to 300 seconds.
If
Unlimited
is checked for a phase, the instrument will NOT use that phase or additional phases.
First result
The interval in seconds between the start of an assay and when the
First Result
displays can be
set manually. This assures that results from shorter, less accurate test times are not displayed.
First test
The time can be set manually before the
First Test
displays and records. Longer test times are
more accurate.
First Test
settings must always be as long as, or longer, than
First Result
times.
Results of the first test are compared to the Grade Library and, if a match is found, the grade ID is
displayed.
Activate beep
This checkbox determines if an audible alarm sounds when a grade match is found.
Trigger
options
Manual Trigger Active
– The instrument operates while the trigger is pressed or until safety
criteria are exceeded.
Auto Trigger Active
– The measurement begins when the trigger is pressed and released. Analysis
time is controlled by defined phase durations. The measurement can be stopped at any time by
pressing and releasing the trigger again. This option is not available in all markets.