4. APPLICATION Button
Description
The
APPLICATION
Button accesses application options. The intended use of the instrument
determines which option to select.
Possible
options
Available applications based on the purchased configuration can include –
Standard alloys.
Restricted materials.
Mining.
Precious metals.
Other custom applications.
Selecting
application
To select an application –
Step
Action
Result
1
From the
Ready to Test
screen,
tap
APPLICATION
.
The
Select Analysis Type
screen is displayed.
2
Tap an application name.
The application name is highlighted.
3
Tap
OK
.
The
Ready to Test
screen is displayed with the
name of the selected application on the
APPLICATION
button.
Spectro-
meter mode
Applications have a preset
voltage, current, and filter that cannot be changed.
Spectrometer
Mode
allows those settings to be adjusted within limits. Spectra are recorded, but data are not
analyzed.
Settings
For each application, settings can be defined. Settings include –
Method
Display options
Phase durations (measurement times)
Manual or automatic trigger
Whenever the application is changed, settings are changed to those defined for that application.