S1 TITAN, T
RACER
5, and CTX User Manual
030.0112.04.0
14 of 67
1.3.
General Care and Maintenance
Low
maintenance
The instrument, when used properly, should require very little maintenance beyond –
Battery recharging. See
Power Sources
, page 15.
Window replacement. See below.
Quality checking. See
Quality Check
, page 61.
Touch screen
The touch screen uses sensitive electronics and should be cleaned regularly using a soft, lint-free
or microfiber cloth.
Do not –
Use cleaners as they may damage the screen.
Use compressed air.
Leave the analyzer where the touch screen is exposed to direct sunlight.
Cleaning
exterior
If the exterior of the analyzer becomes dirty, wipe it gently with a damp, lint-free cloth.
Avoid
harsh solvents and compressed air.
1.4.
Changing Windows
Description
The window in front of the detector is delicate, only 4 µm thick Prolene™, and periodically
requires replacement.
Never touch or use compressed air on the detector. The detector window is 8µm Be, fragile, and
expensive.
Detector