16
en
12. Technical Service
►
For service and repairs, please contact your national B. Braun/Aesculap
agency.
Modifications carried out on medical technical equipment may result in
loss of guarantee/warranty rights and forfeiture of applicable licenses.
Service addresses
Aesculap Technischer Service
Am Aesculap-Platz
78532 Tuttlingen / Germany
Phone:
+49 7461 95-1601
Fax:
+49 7461 14-939
E-Mail:
Or in the US:
Aesculap Implant Systems LLC
Attn. Aesculap Technical Services
615 Lambert Pointe Drive
Hazelwood
MO, 63042
Aesculap Repair Hotline
Phone:
+1 800 214-3392
Fax:
+1 314 895-4420
Other service addresses can be obtained from the address indicated above.
13. Accessories/Replacement Parts
Power cord
14. Technical Data
14.1
Classification acc. to Directive 93/42/EEC
14.2
Performance data, information about standards
14.2.1 OrthoPilot
®
4 panel PC FS112
14.2.2 Power supply for OrthoPilot
®
4 Panel-PC FS112830 (Perfor-
mance specifications for use with FS112)
WARNING
Risk of injury and/or malfunction!
►
Do not perform any servicing or maintenance
work while the product is being used on a
patient.
►
Do not modify the product.
Art. no.
Designation
FS112830
Power supply for the OrthoPilot
®
4 panel PC
FS112802
OrthoPilot® USB security keys
Art. no.
Designation
FS095
Angled power cable, Switzerland
FS096
Angled power cable, Euro plug
FS097
Angled power cable, UK and Ireland
FS098
Angled power cable, Japan/USA/Canada
Art. no.
Designation
Class
FS112
OrthoPilot
®
4 panel PC
I
FS112830
Power supply for the OrthoPilot
®
4
panel PC
I
Protection class acc. to
IEC/DIN EN 61140
Protection class 1
Input voltage
24 V DC
Current consumption
(max. load)
2.5 A
USV battery
Secondary Li-Ion battery
14.4 V DC; 1 750 mAh
Weight
8.5 kg (
–
5 %)
Dimensions (L x W x H)
450 mm x 388 mm x 100 mm (
–
5 %)
Radio technology
Type:
Bluetooth Classic / Low Energy
Frequency range:
2,402 - 2,480 GHz, ISM
band
RF output power - Classic Bluetooth:
Class 1, Max. 13 dBm
RF output power - Bluetooth low energy:
Max. 8 dBm
Conforming to standard
IEC/ DIN EN 60601-1
EMC
IEC/ DIN EN 60601-1-2, CISPR11 Class A
Radio Standards
ETSI EN 300 328
ETSI EN 301 489-1
ETSI EN 301 489-17
IEC/DIN EN 62311
MTBF (Backlight)
30 000 hours
Protection class acc. to
IEC/DIN EN 61140
Protection class 1
Mains voltage ranges
100 V~ – 240 V~ / (
±
10 %)
Current consumption
1.4 A – 0.5 A
Frequency
50 Hz/ 60 Hz (
±
5 %)
Output voltage
24 V DC (
±
5 %)
Output current
2.5 A
Operational mode
Continuous operation
Weight
0.75 kg (
±
2 %)
Dimensions (L x W x H)
180 mm x 66 mm x 41 mm (
–
1 mm)
Conforming to standard
IEC/ DIN EN 60601-1
EMC
IEC/ DIN EN 60601-1-2, CISPR11 Class A
MTBF (power supply)
40 000 hours