Disassembly / Assembly
3
Infusomat® Space 6.0
3 - 53
EN
6. Unpack the zip file with the new tube data (available on re-
quest) into the following directory
C:\B.Braun Space\HiBaSeD_8.0.1_1\dat\Infusomat.
Alternatively: Copy the unpacked folders “default” and “Dis-
posables” to that same directory.
7. Start “HiBaSeD 8.0.1_1_neue Schlauchdaten” by double-
clicking on the link.
8. Ontab “Service” > “Calibration“, press button “Modify”.
9. Only choose “Disposable article” from the data selection.
10. Start calibration with “OK”.
The new tube data are imported.
11. In menu
Line selection
check whether the new disposable ar-
ticle table exists in the device.
Fig.: 3 - 61
Fig.: 3 - 62
Fig.: 3 - 63
Fig.: 3 - 64