Infusomat® Space 6.0
5 - 1
EN
Index h
(Master – to be added to the documentation)
M686 70 01 05 F04 / 38918838
Sheet 1 of 3
5 -
Technical Safety Check (TSC)
Infusomat® Space
Checklist for Technical Safety Check – Every 24 Months
User
Unit: Infusion Pump Infusomat® Space
Manufacturer: B. Braun Melsungen AG
Observe the Service Manual and the instructions for use. All measured values are to be
documented. Accessories used should be included in testing. Make exclusive use of cali-
brated measuring equipment.
Year of Procurement
Article No.
Unit No.
Stock No.
Visual Inspection
Electrical Safety
According to IEC 60601-1 or IEC 62353
Functional Inspection
❒
Infusomat® Space:
Cleanliness, completeness, damage and
defects affecting safety, damage and
readability of the labels.
Particularly:
❒
Connectors „P2“ and „P3“
❒
Safety clamp
❒
Pressure sensor, downstream
(membrane)
❒
Pressure sensor, upstream (mem-
brane)
❒
Pump membrane
❒
Pump slide
❒
Coding of pump slide guide
❒
Pump connecting rod
❒
Air inline sensor (free from grease
or ultrasonic gel, free from cracks)
❒
Drop sensor (lens)
❒
Accessories:
Cleanliness, completeness, damage and
faults affecting safety, damage and
readability of the label.
❒
Check the unit and the accessories for
compatibility
The patient and housing leakage current of
the Infusomat® Space is caused exclusively
by the operating voltage supply (Power Sup-
ply SP or SpaceStation).
The Technical Safety Checks of the power
supply SP (drawing No. M001321005F04) or
of the SpaceStation (drawing No.
M690000046F04) serve to check whether
both limit values are met.
❒
Locking with second unit
❒
Battery compartment cover
Determine age of battery
❒
If older than 8 years: recommend re-
placement of battery or perform bat-
tery check (see
)
Switch on unit with power supply:
❒
Self-test
❒
Indicator lamps (LEDs)
❒
Audible alarm
❒
Visual alarm
❒
Status display
❒
Lighting
Operation
❒
Opening and closing mechanism of the
operating unit
❒
Slide clamp releases release lever of the
safety clamp. The Infusomat® Space
Line is squeezed.
❒
Buttons on the operating unit
❒
Infusion
❒
Staff call
5 - 4
Edit
Summary of Contents for 0871 3050
Page 1: ...Infusomat Space Service Manual Version 6 0 English 0 12 0 RX only...
Page 12: ...Contact Persons 0 0 12 Infusomat Space 6 0 EN For your notes...
Page 20: ...System Overview 1 1 8 Infusomat Space 6 0 EN Fig 1 7 Block Diagram Infusomat Space...
Page 30: ...System Overview 1 1 18 Infusomat Space 6 0 EN For your notes...
Page 108: ...Servicing the Unit 4 4 4 Infusomat Space 6 0 EN For your notes...
Page 130: ...Spare Parts List 8 8 6 Infusomat Space 6 0 EN For your notes...
Page 135: ......