Unit Diagnosis / Calibration
2
Infusomat® Space 6.0
2 - 7
EN
CAN error states
Signalling of a CAN error
In case of SW F, a faulty CAN bus is signalled by the blue LED at
the pump flashing. The LED flashes at the following speeds:
-
300 ms on, 300 ms off
-
100 ms on, 100 ms off
-
100 ms on, 100 ms off
The process is then repeated.
The blue LED flashes in the following states:
Behaviour when an alarm occurs
If continuous flashing occurs at the pump, the pump is fully op-
erational with the exception of communication-dependent ther-
apies.
The following functions are no longer available in this state:
-
Signalling of the alarm states at the SpaceCover
-
Data communication with SpaceCom (no documentation for
a PDMS)
-
Proposal data can no longer be received.
-
In case of a defective address assignment, communication
with HiBaSeD is not possible until after plugging into a slot
which is not defective. (In the case of duplicate serial number,
HiBaSeD can contact the pumps in order to carry out calibra-
tion, see Section 6)
-
Future therapies, such as take-over, patient data, bar code
and remote control function, will not be available either.
Alarm
Cause
Fault Clearance
1
CAN lists for 2 seconds after pushing into
a Station
Check that the pump is pushed in cor-
rectly
2
CAN_Lists because of duplicate address
assignment
Duplicate address assignment
This state remains active until another
device is recognised with its own slot ad-
dress.
3
CAN dead because of duplicate INR
Duplicate INR
Eliminate the state by calibrating with
HiBaSeD. The device is still ready for op-
eration but without CAN communication.
Table 2 - 4
Summary of Contents for 0871 3050
Page 1: ...Infusomat Space Service Manual Version 6 0 English 0 12 0 RX only...
Page 12: ...Contact Persons 0 0 12 Infusomat Space 6 0 EN For your notes...
Page 20: ...System Overview 1 1 8 Infusomat Space 6 0 EN Fig 1 7 Block Diagram Infusomat Space...
Page 30: ...System Overview 1 1 18 Infusomat Space 6 0 EN For your notes...
Page 108: ...Servicing the Unit 4 4 4 Infusomat Space 6 0 EN For your notes...
Page 130: ...Spare Parts List 8 8 6 Infusomat Space 6 0 EN For your notes...
Page 135: ......