Procedural Instructions on the TSC
6
6 - 6
Infusomat® Space 6.0
EN
Test Setup
Perform test setup with the subassemblies listed below, please see
also
Note
The filling level in the container must be approx. 50 cm above the
middle of the unit opening for the Infusomat® Space Line.
Infusomat® Space Line, standard type
(new, unused, can be used for the complete TSC incl. func-
tional check)
(1 piece)
Infusion bag or bottle, min. 100 ml
(1 piece)
Three-way valve (2 pieces for measurement with
electronic pressure gauge)
10 ml syringe (air buffer for measurement with
electronic pressure gauge)
(syringe drawn up to 10 ml and piston fixed mechanically)
(1 piece)
1 ml syringe for air bubble injection (1 piece)
Electronic pressure gauge with peak value recognition (1 piece)
Graduated cylinder 25 ml, ±0.4 ml (1 piece)
Pressure Cut-Off, Electronic
The Space service connector must not be connected to the device.
The optional drop sensor must not be connected.
1. Insert the Infusomat® Space Line of the test setup (
2. Enter a delivery rate according to the TSC and a volume of
250 ml.
3. Select pressure stage according to the TSC.
Fig.: 6 - 5
Legend of fig. 6 - 5:
Item Designation
1 Electronic pressure gauge accuracy class 1
2 10 ml syringe (piston fixed)
3 Infusion bottle
4 Infusomat® Space Line, standard type
5 Distance between liquid level and middle Infusomat® Space
Line
6 Three-way valve 1
7 Three-way valve 2
8 Graduated cylinder
1
2
appr. 50 cm
Summary of Contents for 0871 3050
Page 1: ...Infusomat Space Service Manual Version 6 0 English 0 12 0 RX only...
Page 12: ...Contact Persons 0 0 12 Infusomat Space 6 0 EN For your notes...
Page 20: ...System Overview 1 1 8 Infusomat Space 6 0 EN Fig 1 7 Block Diagram Infusomat Space...
Page 30: ...System Overview 1 1 18 Infusomat Space 6 0 EN For your notes...
Page 108: ...Servicing the Unit 4 4 4 Infusomat Space 6 0 EN For your notes...
Page 130: ...Spare Parts List 8 8 6 Infusomat Space 6 0 EN For your notes...
Page 135: ......