Unit Diagnosis / Calibration
2
2 - 12
Infusomat® Space 6.0
EN
4. If requested by the device, enter the date and time.
5. Select a language from the list in the device and confirm your
selection by using the "OK" key on the device. The "Worker ID"
window opens.
Note
If you have not been allocated a user number (Worker ID), enter
"0001".
6. Enter your user number in the window “Worker ID” as well as
the six-digit serial number of the device, if necessary.
7. Confirm the entry with “OK“. The “Data selection” group box
is now activated.
Note
If HiBaSeD could not clearly read the device serial number, the
number on the rating plate must be entered.
Fig.: 2 - 2
Fig.: 2 - 3
Fig.: 2 - 4