Disassembly / Assembly
3
3 - 32
Infusomat® Space 6.0
EN
3. Unscrew 6 screws (
) and remove the seal
) from the operating unit rear
panel (metal front sheet, Outsert, Version II) (
).
Note
The screws of the keyboard must not be loosened. A uniform pres-
sure point of all keys is only guaranteed when a special tool is
used for assembly.
4. Remove the operating unit (
)from the rear
panel (metal front sheet, Outsert, Version II).
Note
Please pay attention to the ground connection(
) on the keyboard PCB during disassembly.
Fig.: 3 - 30
Legend of fig. 3 - 30:
Item Designation
1 Countersunk screw EJOT 20x9 WN 5454 TORX plus 6IP
(tighten finger tight)
2 Seal washer PA 2.2x0.3
3 Operating unit rear panel (metal front sheet, Outsert,
Version II)
4 Ground connection
5 Operating unit with LC display