Disassembly / Assembly
3
Infusomat® Space 6.0
3 - 57
EN
Pressure switch-off, electronic:
(see TSC for limit values)
Delivery rate: 250 ml/h
Testing without drop sensor
Infusomat® Space Line, standard type
❒
Upstream
❒
Alarm
❒
Downstream
❒
Pressure stage 2
❒
Pressure stage 5
❒
Pressure stage 9
Pressure limitation, mechanical
(see TSC for limit values)
❒
Pmax
❒
Pmin
Safety clamp
(see TSC for limit values)
❒
Pmin
Flow accuracy
(see TSC for limit values)
Room temperature 20 … 25°C, delivery rate: 250 ml/h,
measuring quantity: 25 ml
❒
Deviation within tolerance
Drop sensor (optional)
Delivery rate: 250 ml/h
❒
Drop sensor alarm passage
❒
Drop sensor alarm closure
Visual Inspection
Electrical Safety
According to IEC 60601-1 or IEC 62353
Functional Inspection
(Part 2 of 3)