![Bosch CCT 3 Series Test Instructions Download Page 28](http://html.mh-extra.com/html/bosch/cct-3-series/cct-3-series_test-instructions_50401028.webp)
en
1 690 386 058
2018-01-30
|
Robert Bosch GmbH
28 | SCT / CCT / EL | User instructions
en
2.1.2
Test board 1 690 382 354
1
4512001-04_shd
Abb. 3: Test board 1 690 382 354
1 Zero point on scale
2.1.3
Laser module 1 690 381 124
4512001-03_shd
Abb. 4: Laser module 1 690 381 124
i
The customer's laser module is identified as "P-
Assist S5" in the following. The Bosch Customer Ser-
vice laser module is identified as "reference laser" in
the following. The reference laser is needed to verify
the functionality of the P-Assist S5.
2.2
Test requirements
To test an SCT / CCT / EL, it is necessary to have a free
wall on which to fasten the test boards 1 690 382 354
by means of double-sided adhesive tape. If a free wall is
not available, a different surface can be used, e. g. the
column of a lift.
2.3
Definition of the angle being checked
1
2
3
Abb. 5: Test angle
1 Yaw angle
2 Roll angle
3 Pitch angle
2.4
Requirements
Setting
Specification
Pitch angle
+/- 0.5°
Roll angle
+/- 0.3°
Rotation around vertical axis for laser mo-
dule
+/- 0.15°
Alignment of the measurement boards
+/- 1.5 mm
Lateral positioning of the test boards
+/- 4 mm
Height adjustment
+/- 800 - 1500 mm
Accuracy of the height adjustment
+/- 5 mm
Yaw angle
+/- 0.2°
Tab. 1: