MS-301 Multi parameter Water Quality Sensor
Web:www.boquinstrument.com
2470
electric voltage of the
fifth point standard
solution calibration of
electrode 3
float
read
2472
temperature of the fifth
point standard solution
calibration of electrode 3
float
read
22
value of electrode 4
float
read
command 03
2500
electrode voltage of
electrode 4
float
read
2502
factor of electrode 4
float
read/write
Default:1
2504
deviation of electrode 4
float
read/write
Default:0
2506
type of electrode 4
float
read/write
The types of electrode are
represented by floating point
number of 1, 2, 3, 4, 5, and 6.
1 represents ORP;
2 represents ammonia
nitrogen;
3 represents Nitrate ion; 4
represents chloride ion;
And 5 represents fluoride ion.
2508
calibration points of
electrode 4
UINT16
read/write
Support the command o06.The
written values of 1, 2, 3, 4, and
5 respectively represent single
point calibration to 5 point
calibration.
2515
stability
float
read
Represents the stability of the
electrode voltage (used during
calibration, the value on 0.9
represents the stability)