MS-301 Multi parameter Water Quality Sensor
Web:www.boquinstrument.com
2 represents ammonia
nitrogen;
3 represents Nitrate ion; 4
represents chloride ion;
And 5 represents fluoride ion.
2308
calibration points of
electrode 2
UINT16
read/write
Support the command of 06.
The written values of 1, 2, 3,
4, and 5 respectively represent
single point calibration to 5
point calibration.
2315
stability
float
read
Represents the stability of the
electrode voltage (used during
calibration, the value on 0.9
represents the stability)
2334
first point calibration of
electrode 2
float
write
input the standard solution
value
2336
the first point standard
solution value of
electrode 2
float
read
2338
electric voltage of the first
point standard solution
calibration of electrode 2
float
read
2340
temperature of the first
point standard solution
calibration of electrode 2
float
read
2342
second point calibration
of electrode 2
float
write
input the standard solution
value
2344
the second point standard
solution value of
electrode 2
float
read