MS-301 Multi parameter Water Quality Sensor
Web:www.boquinstrument.com
6.2.6 Turbidity calibration
Get into the interface of turbidity calibration, and press the key of “OK” at the site of “wiping before
calibration”. Press the key of “Enter” to perform cleaning procedure for one time by scratching, and
eliminate the influence of bubbles on turbidity calibration process.
Choose first point/second point turbidity calibration according to actual requirement to make integrated
multi parameter water quality sensor immerge into standard solution, and at the same time, input
corresponding standard solution value in the square frame of the hand manipulator. Then, choose the key
of “OK” and press "Enter" for calibration and confirm the calibration status according to the prompt of
calibration status. When the calibration is indicated to be successful, the calibration at this point is
complete.
6.2.7 Chlorophyll calibration
Get into the interface of chlorophyll calibration, and make integrated multi parameter water quality sensor
immerge into distilled water or rinsing. Press the key of “OK” at the site of “wiping before calibration”.
Press the key of “Enter” to perform cleaning procedure for one time by scratching, and eliminate the
influence of bubbles on chlorophyll calibration process. Press the key of “OK” at the site of “zero