MS-301 Multi parameter Water Quality Sensor
Web:www.boquinstrument.com
calibration container. Carefully immerse the Sensor at the end of the main unit in the solution. Please
ensure that the front end of the Chloride Ion Sensor is immersed in the solution at least 2 cm and enter the
standard value. When the value in the software interface is stable, click “3 point Calibration”. Rinse and
dry the sensor and calibration container with water after completion.
You can directly select the input value in the correction parameter box and click “factor” and “deviation”
to calibrate the factor and deviation value.
6.2 Calibrate the sensor by hand manipulator
Choose “calibration” in the main menu page, and press the key of “Enter” to confirm. Then, you can get
into the calibration interface of selective parameters. The parameters you choose must be the Sensors that
have been connected as showed in the diagram below.
Please check the details of all the specific calibration steps of sensors below in
the Chapter 6.1
6.2.1 Temperature calibration
The new product of temperature sensor has been calibrated, and the default of the temperature source is Pt
1000.It is not necessary to calibrate again.