MS-301 Multi parameter Water Quality Sensor
Web:www.boquinstrument.com
calibration point to 322
calibration register of standard
solution of blue-green algae.
324
calibration point setting
of blue-green algae
float
write
Set the calibration point of
blue-green algae. (180
represents distinct calibration
data, 181 represents the first
point, 182 represents the
second point, 183 represents
the third point, and 184
represents the fourth point).
530
interval of scratching
float
read/write
unit: min
532
manual scratching
float
Write
1 represents scratching
70
year
float
read/write
72
month
float
read/write
74
day
float
read/write
76
hour
float
read/write
78
minute
float
read/write
80
second
float
read/write
9995
slave baud rate
float
read/write
The range of setting for baud
rate includes 4800, 9600,
19200, 38400, and 115200.
And the floating point number
of 1, 2, 3, 4, and 5 will
represent corresponding baud
rate.
9997
slave address
unsigned
read/write
Write using command 06,
range:1-255
16
value of ion electrode 1
float
read
command 03
2200
electrode voltage of
float
read