MS-301 Multi parameter Water Quality Sensor
Web:www.boquinstrument.com
calibration” in the interface, and press the key of “Enter”.
Make integrated multi parameter water quality sensor immerge into Rhodamine B standard solution, and
input corresponding sample value of water at the site of “sample calibration”. Press the key of “OK”, and
press the key of “Enter”.
6.2.8 Blue-green algae calibration
Get into the interface of blue-green algae calibration, and make integrated multi parameter water quality
sensor immerge into distilled water or rinsing. Press the key of “OK” at the site of “wiping before
calibration”. Press the key of “Enter” to perform cleaning procedure for one time by scratching, and
eliminate the influence of bubbles on blue-green algae calibration process. Press the key of “OK” at the
site of “zero calibration” in the interface, and press the key of “Enter”.
Make integrated multi parameter water quality sensor immerge into Rhodamine B standard solution, and
input corresponding sample value of water at the site of “sample calibration”. Press the key of “OK”, and
press the key of “Enter”.
6.2.9 Fluoride ion calibration
Get into the interface fluoride ion calibration, and press the key of “calibration”. Press the key of “Enter”