MS-301 Multi parameter Water Quality Sensor
Web:www.boquinstrument.com
value
334
conductivity Sensor
calibration
float
read /write
Input the standard liquid value
6
dissolved oxygen value
float
read
0-20 mg/L
1606
dissolved oxygen factor
float
read/write
Default:1
1506
dissolved oxygen
deviation
float
read/write
Default:0
904
dissolved oxygen
calibration
float
write
Send 2, and wait for at least 20
seconds. And then, send 3 to
complete the air calibration.
430
salinity compensation
model
float
read/write
0: no compensation
1: manual compensation
2:salinity sensor compensation
432
salinity compensation
value
float
read/write
Register 430 salinity
compensation mode is manual
compensation mode, this value
is used to compensate
8
turbidity value
float
read
0.01~1000NTU
1010
turbidity AD value
float
read
1608
turbidity factor
float
read/write
Default:1
1508
turbidity deviation value
float
read/write
Default:0
1100
turbidity calibration
progress value
float
read
0 represents successful
calibration
1 represents wrong calibration
2 represents entering the
calibration model
20 represents it is in the
process of calibration
255 represents the lock of
calibration has been unfolded,