MS-301 Multi parameter Water Quality Sensor
Web:www.boquinstrument.com
Make sure the depth sensor is clean and dry before calibration. For depth sensor calibration, it is mainly
for atmospheric pressure calibration, so make sure the depth sensor is in the air instead of being immersed
in any container.
The specific calibration steps are as follows:
Try to ensure that the sensor is dry before calibration. When the atmospheric pressure value in the
calibration interface is stable, click “Pressure Calibration” in the “Pressure Calibration” box. The system
will automatically save the data to complete the calibration. Due to the special nature of the depth sensor,
it is recommended that customers need to calibrate frequently to maintain data accuracy.
You can also directly enter the value in the pressure correction parameter box, click “factor”, “deviation”
to calibrate the factor and deviation value, or directly input the value in the depth correction parameter
box, click “factor” and “deviation”. Factor and deviated values are calibrated.
6.1.8 Optical Sensor Calibration
6.1.8.1Turbidity Sensor Calibration
From the Calibration menu, select the Turbidity option to enter the turbidity calibration procedure as
shown below, and you are ready to begin calibration.