Integrated Multi-parameter Sensor Operation manual
79
represented by floating point
number of 1, 2, 3, 4, 5, and 6.
1 represents ORP;
2 represents ammonia
nitrogen;
3 represents Nitrate ion; 4
represents chloride ion;
And 5 represents fluoride ion.
2408
calibration points of
electrode 3
UINT16
read/write
Support the command of
06.
The written values of 1, 2,
3, 4, and 5 respectively
represent single point
calibration to 5 point
calibration.
2415
stability
float
read
Represents the stability of the
electrode voltage (used during
calibration, the value on 0.9
represents the stability)
2434
first point calibration of
electrode 3
float
write
input the standard solution
value
2436
the first point standard
solution value of
electrode 3
float
read
2438
electric voltage of the first
point standard solution
calibration of electrode 3
float
read
2440
temperature of the first
point standard solution
float
read