Integrated Multi-parameter Sensor Operation manual
70
255 represents the lock of
calibration has been unfolded,
and it is well ready for the
operation of calibration
906
turbidity calibration
model
float
write
enter 1 to access the
calibration model
1014
turbidity low point
calibration
float
write
Enter 1 into the turbidity
calibration mode of register
address 906 first, so that the
turbidity sensor enters into the
calibration mode, and then
enter the actual value of the
standard liquid
1016
turbidity high point
calibration
float
write
Enter 1 into the turbidity
calibration mode of register
address 906 first, so that the
turbidity sensor enters into the
calibration mode, and then
enter the actual value of the
standard liquid
10
depth value
float
read
0~61m
40
pressure value
float
read
common electrode: ≤0.3Mpa
ion electrode: ≤0.1 Mpa
1604
depth factor
float
read/write
Default
:
1
1504
depth deviation value
float
read/write
Default
:
0
130
atmosphere pressure
calibration
float
read/write
enter any value can realize the
calibration
12
chlorophyll value
float
read
0.01~500ug/L
1020
chlorophyll AD value
float
read