Integrated Multi-parameter Sensor Operation manual
73
rate.
9997
slave address
unsigned
read/write
Write using command 06,
range:1-255
16
value of ion electrode 1
float
read
command 03
2200
electrode voltage of
electrode 1
float
read
2202
factor of electrode 1
float
read/write
Default
:
1
2204
deviation of electrode 1
float
read/write
Default
:
0
2206
type of electrode 1
float
read/write
The types of electrode are
represented by floating point
number of 1, 2, 3, 4, 5, and 6.
1 represents ORP;
2 represents ammonia
nitrogen;
3 represents Nitrate ion; 4
represents chloride ion;
And 5 represents fluoride ion.
2208
calibration points of
electrode 1
unsigned
read/write
Support the command of 03
and 16. The written values of
1, 2, 3, 4, and 5 respectively
represent single point
calibration to 5 point
calibration.
2215
stability
float
read
Represents the stability of the
electrode voltage (used during
calibration, the value on 0.9
represents the stability)
2234
first point calibration of
electrode 1
float
write
input the standard solution
value