18
7.2
Program Settings
Press the Program softkey in the Setting and Measurement page to enter Program mode.
There are three modes for the user to choose from to simulate Power Line Disturbances.
They are:
Step
Output settings step-up or step-down based on user criteria.
List
Change the output sequentially by individual settings included in a list.
Pulse
Trigger pulse signals periodically.
7.2.1
Step Mode
Step Mode has 10 parameters that can be set by the user. The following table describes
each parameter with its range and description.
Parameters
Range
Description
Volts (rms)
0 to 300.0 Vrms
Initial AC output voltage
dVac
0 to ±300.0 Vrms
In/decrement of VAC per
step
Volts (DC)
-424.0 to 424.0 V
Initial DC output voltage
dVdc
-424.0 to 424.0 V
Increment/decrement of
VDC per step
Freq.
43 to 1200.0 Hz
Initial output frequency
dF
±43 to 1200.0 Hz
Increment/decrement of
FREQ. per step
Set Time
0 to 100000 ms
Set the time interval of
one step
Count
1 to 99
Set how many steps to
run
Sync Source
Immed, Phase
Select the mode of
output transient phase
Phase (deg)
0.0
゜
to 359.7
゜
Set the angle of output
transient
Table 2 - Step Mode Parameters