20
2.
Use the arrow keys (
▲
▼
◄
►
) to move the cursor to the wanted parameter of
the STEP setting page.
3.
Press the ENTER key to adjust the parameter. Then press the ENTER key again to
change its value.
4.
When finished editing all the parameters, press On/Off key to activate the Step
mode.
Using an oscilloscope you should observe a wavfore as shown in Figure 16 - Oscilloscope
view of Step waveform example.
Figure 16 - Oscilloscope view of Step waveform example
The instrument will continuosly output the final waveform of Step mode until the On/Off
key is pressed.
7.2.2
List Mode
List mode is the only mode with two pages of parameters. The first page of parameters
is common to all the Second page parameters. Only one first page is needed per list. The
first page, Table 4 - List Parameters, First Page is displayed when the List softkey is
pressed and the second page, Table 5 - List Config Parameters, Second Page is displayed
when the List Config softkey is pressed.
Parameters
Range
Description
List
0 to 9
Index of LIST
Infinite
ON or OFF
ON: Infinite output
OFF: Based on Repeat value
Repeat
0 to 99
Repeat sequence
Base
Time or Cycle
Select the unit for timing the
running of LIST
Sync Source
Immed or Phase Select the mode of output
transient phase
Phase (deg)
0.0 to 359.7
Set the angle of output transient
Table 4 - List Parameters, First Page