11
current required to operate the instrument.
2.
Using the power cable provided, identify the end that plugs into the instrument.
Before disassembling the black hooded connector, make a note of how the unit is
assembled from the factory so it can be correctly reassembled.
3.
Remove the 4 screws from the bottom of the hooded connector as shown and place
the upper and the upper and lower hood pieces and screws to the side.
The black, white, and green wires should now be clearly visible. Use a screwdriver to
unscrew and release the white wire only.
Add a band of blue electrical tape or heat shrink to each end of the white wire.
1.
Return the white wire into the green connecter and tighten the screw. Verify all
three wires are secured in the connector.
2.
Closely inspect each of the wires at the enter point of the green connector to
verify none of the copper strands are bridging over to the wrong terminal
causing a short.
3.
Carefully reassemble the hood so the Latch Fingers and Caution Label are facing