BT-750 Operator Manual 56
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
▣
NIBP measurement
Problem
Solution
NIBP measurement is not done
properly
- Reset the power
- Contact our customer service center or repair center.
NIBP measurement is failed
-Make sure if the patient uses proper cuff.
-Make sure if the posture of the patient and the portion to
be measured are suitable.
-Contact our customer service center or the repair center.
▣
IBP measurement
Problem
Solution
Blood
pressure
measurement is unstable.
Pour out the contents of the tube carefully and remove the
bubble, then shorten the tubing length.
Severe noise or bias in the
waveform.
Do not shake or touch the connection area of the IBP sensor.
▣
EtCO2 measurement
Problem
Solution
FILTERLINE NOT
CONNECTION
The filter line is not connected to the main body. Connect filter
line disconnected the filter line.
EtCO2, InCO2 are
Blank
Contact the agency.
EtCO2 waveform and
numeric are not
displayed
Check if the sampling line and filter is obstructed, and clean it up
if possible. It should be changed if necessary. To remove this
message, disintegrate the filter line and then reconnect it.
▣
Problems in the Printer measurement (Optional)
Problem
Solution
No printing.
1.
Check if the message “PRINTER” is displayed on the screen.
2.
Check if the power LED lights up on the printer.
3.
Check if the paper is set normally.
4.
Check if the printer door is closed.
Noise during printing.
1.
Check if the roller is contaminated with foreign materials or
if the gear is out of position.
Check if the problem is due to a paper jam (change the paper)
Please contact our customer service center or repair center if there is any defects or inquiry.