BT-750 Operator Manual 28
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
2) Alarm icon
The alarm icon appears on the information window of the lower screen..
Function
Message
Description
SpO
2
STAND BY
Probe is connected but the finger is out of probe.
CHECK PROBE
Finger probe is out of probes
ECG
CHECK SENSOR
Sensor is out of body or cable is out of the monitor
IBP
Temperature
EtCO2
NIBP
CUFF HOSE
Failure of pressurization for definite time
NO PULSE
Fail to detect pulse rate
REDA FAIL
Error in measurement values for moving
ALARM
LOW
Refer to Alarm Generating Chart
MIDIULM
HIGH
[Alarm Generating Chart]
Message
Generating Cause
HIGH
When pulse rate is not detected (“0” BPM)
MEDIUN
Out of numeric value
LOW
1. SP02 PROBE ERROR 2. LOW BATTERY
3. Fail to measure NIBP or Over Pressure
4. Lead Fault during ECG measurement
5. Finger is out of SpO2 probe during SpO2 measurement
6. Fault during Temp measurement 7. Fault during IBP measurement
8. PACEMAKER Detect
Alarming display is illuminated in every alarming
.
It is changed to the below chart,
in every moment pressing alarming button ,when alarm generates.
NO
DISPLAY
SOUND
REMARK
1
Normal alarming
Default
2
Holding alarm for 60 sec
.
Pressing alarm button once
3
Alarm sound off for 120 sec.
Pressing alarm button twice
4
Alarm off
Refer to the chapter 14
Please press the alarm button ‘once’ in low level alarm to cancel it.
Caution
Messages generate and indicators are illuminated, regardless of
alarming or not.
3) Setting Up Alarm Hold
After alarm generates, user can hold audio alarm and set up the time to 1min/2min/normal.
If each event which generates alarming within setting time is not cancelled, audio alarm will
generate again.