BT-750 Operator Manual 59
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
IBP
Channel
Method Type
Pulse rate range
Pulse rate accuracy
Pressure Measurement range
Parameter Displayed
Input Impedance
Transducer Driving Voltage
Transducer input Sensitivity
Transducer Volume Displacement
Zero Calibration Range
Zero Calibration Accuracy
Frequency Characteristics
Pressure Display Accuracy
Scale
Display Sweep Speed
2 Channel
Pressure Transducer Method
20 to 250 bpm
±1 % or ±1 bpm
0 to +300 mmHg
P1, ABP
P2, CVP, PAP, LAP
More than 1 M Ω
DC 5 V
5
㎶
/V/mmHg
0.1
㎣
/mmHg
±100 mmHg
Less than ±1 mmHg
Dc to25
㎐
Monitor: Less than ±3 mmHg
P1 0~50, 0~100, 0~200, 0~300,
P2 0~25, 0~50, 0~100, 0~200, 0~300
6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s
Capnography
CO2
Parameter Displayed
Method type
Rise Time
Warm Up Time
Sweep Speed
CO2 Measurement range
CO2 Accuracy
EtCO2, InCO2
Non-dispersive infrared
Single beam optics, dual Wavelength,
no moving parts
less than 60ms
2 minutes maximum
6.25 mm/s, 12.5 mm/sec, 25.0 mm/sec
0 to 150 mmHg
0 to 20 kPa
0 to 19.7 %
0 to 40mmHg ±2mmHg
41 to 70mmHg ±5% of reading
71 to 100mmHg ±8% of reading
101 to 150mmHg ±10% of reading