BT-750 Operator Manual 52
P/N : OPM(BT-750)EN Bistos Co.,Ltd. Rev.02(2013.04.12)
11.2 How to measure Sidestream CO2
1) Attach the gas/tube to BT-750 Sidestream port and the other side attach to hospital gas
device
2) Respiration numeric is automatically changed when EtCO2 is measured.
3) EtCO2 will be displayed on BT-750 screen after 20 seconds when it detect proper input.
4) EtCO2 waveform and data is displayed. If waveform is not displayed on the screen, change
setting menu “Menu → EtCO2 → EtCO2 → Enable
5) You can change EtCO2 waveform when you need.
Caution
The watertrap is disposable and should only be used for a single
patient. Do not reuse the watertrap for another patient.
Caution
The cannula is disposable and should only be used for a single
patient. Do not reuse the cannula for another patient.
Caution
If oxygen is being delivered while using Sidestream CO2, be sure to
use a CO2 Sampling and O2 Delivery cannula. Using a different type
of cannula could obstruct oxygen delivery.
Warning
Assembly, extension, repair should be carried out by authorized
person from Bistos Technology.